![]() |
NKF
KDOQI GUIDELINES 2000
|
||||||
![]() |
![]() |
![]() |
![]() |
||||
V. Hemodialysis Dose Troubleshooting
Inadequate Delivery of Hemodialysis (Opinion)
If the delivered Kt/V falls below 1.2, or the URR declines to <65%, on a single determination, at least one of the following actions should be performed:
1. Investigate potential errors in the delivery of the prescribed hemodialysis dose (see Table I-5);
2. Empirically increase the prescribed dose of hemodialysis; and/or
3. Suspend use of the reprocessed hollow fiber hemodialyzer.
The impact of these corrective interventions should be followed by performing more frequent measurements of Kt/V or URR.
Rationale Every effort should be made to ensure that patients receive adequate hemodialysis at each treatment (see Guideline 4: Minimum Delivered Dose of Hemodialysis). Whenever the delivered dose of dialysis falls below the minimum acceptable level (Kt/V = 1.2 or URR = 65%), a systematic investigation should be initiated promptly to identify why the delivered dose has declined to that level. The HD Adequacy Work Group developed an algorithm (see Table I-5) that can be used to identify the causes of inadequate delivery of hemodialysis and to implement corrective actions. The algorithm is based on the following assumptions:
1. It is unrealistic to expect dialysis care teams to immediately identify and correct problems that result in inadequate hemodialysis. Some delay in identifying a problem typically will occur. For example, with routine blood sampling, there is usually a delay of several days to 1 week between blood sampling and the delivery of the lab report that signals a problem. Over that interval, up to three hemodialysis treatments will have occurred. There will be additional delays as the potential causes of the problem are investigated. A prompt corrective response will minimize inadequate treatments.
2. More frequent measurements of Kt/V or URR will assist in identifying the problem(s), the underlying cause(s), and the necessary corrective action(s).
3. Hemodialysis units and clinicians should strive to categorize failures and to identify specific causes, eg, procedural errors by staff, equipment problems, deficient reprocessing techniques/systems, patient nonadherence, etc.
4. Increasing the prescribed hemodialysis dose until appropriate minimum targets are reached will improve patient outcomes.
5. Inadequate performance of reprocessed dialyzers may be associated with reductions in the delivered dose of hemodialysis.72,204-206,209
6. Prudent clinical judgment requires that use of a potentially inadequate hemodialyzer be suspended until the specific etiology of the insufficiency of delivered dose can be identified. If other factor(s) explaining the deficiency are identified, the dialyzer can be returned to use.
|
Error analysis algorithm Initial Assessments Elements of hemodialysis treatment* to evaluate/correct immediately. These elements can be assessed noninvasively, using available data, and/or without incurring cost. Clearance (K) less than assumed. Elements of the hemodialysis procedure affecting clearance include: dialyzer permeability (KoA), effective dialyzer surface area, blood flow, and dialysate flow.
Effective hemodialysis treatment time (Td) less than prescribed. Hemodialysis treatment time is the total time at the prescribed blood and dialysate flow rates with the prescribed dialyzer, or is the dialysis time determined to provide an equivalent Kt/V at the prevailing blood and dialysate flow rates for a particular hemodialyzer.
Errors in blood sampling or processing for BUN Concentration.
Secondary Assessments Evaluation of these elements of hemodialysis treatment requires further investigation. These investigations may be invasive, require more staff time or equipment maintenance, and/or are cost generating. Clearance (K) less than assumed.
Effective hemodialysis treatment time (Td) less than prescribed.
Errors in sampling or processing for BUN concentration.
* The hemodialysis care team should note that list is not all-inclusive. However, the elements provided are the most common sources of error. |
|
© 2001 National Kidney Foundation, Inc web version created by cyberNephrologyTM and The Nephron Information Center |